AI -40°C Cold Trap with LS Stainless Steel Traps ETL
AI -40°C Cold Trap with LS Stainless Steel Traps ETL
AI -40°C Cold Trap with LS Stainless Steel Traps ETL
AI -40°C Cold Trap with LS Stainless Steel Traps ETL
AI -40°C Cold Trap with LS Stainless Steel Traps ETL
AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL
  • Load image into Gallery viewer, AI -40°C Cold Trap with LS Stainless Steel Traps ETL

AI -40°C Cold Trap with LS Stainless Steel Traps ETL

Vendor
Across International
Regular price
$4,200.00 USD
Sale price
$4,200.00 USD
Regular price
Sold out
Unit price
per 
Shipping calculated at checkout.

The Ai 2nd Gen -40°C/F T40 cold traps are ETL certified. They have a much smaller footprint, but at the same time, offering you the same cooling power.

In vacuum applications, a cold trap is a device that condenses all vapors except the permanent gases into a liquid or solid. The most common objective is to prevent vapors being evacuated from an experiment from entering a vacuum pump where they would condense and contaminate it. Particularly large cold traps are necessary when removing large amounts of liquid as in freeze drying.

Pumps that use oil either as their working fluid (diffusion pumps), or as their lubricant (mechanical rotary pumps), are often the sources of contamination in vacuum systems. Placing a cold trap at the mouth of such a pump greatly lowers the risk that oil vapors will back stream into the cavity.

Care should be taken when using a cold trap not to condense liquid oxygen (a light blue liquid) into the cold trap. Liquid oxygen is potentially explosive, and this is especially true if the trap has been used to trap solvent. Liquid oxygen can be condensed into a cold trap if a pump has sucked air through the trap when the trap is very cold.

PLEASE USE OUR CONTACT US LINK FOR ADDITIONAL SIZES AND OPTIONS

Ai T40-UL -40°C/°F cold trap features

  • Comes with Lab Society stainless steel trap
  • No dry ice or liquid nitrogen needed
  • All steel housing for durability
  • Small footprint, built-in swivel casters for mobility
  • Plug and play, no assembly required
  • ETL tested to UL and CSA standards

 

Specifications

 Model  T40-UL
 Electrical requirements
 110V or 220V, 60Hz 1-PH
 Compressor  Make: Embraco
 Refrigerant: R404A (220g)
 Cooling capacity:
    @ -10°C: 300W
    @ -20°C: 250W
    @ -30°C: 150W
    @ -40°C: 50W
 Reservoir
 Volume
   1.3 gallon (5 liters)
 Dimensions
   Di
ameter 6 1/4 inches
   Depth: 9 1/2 inches
 Temperature
 -40°C (-40°F)
 Display: digital
 Unit: °C
 Sensor: PT100
 Finish  Exterior: powder coated steel
 Interior: 
polished type 304 stainless steel
 Trap  Material: stainless steel
 Make: Lab Society
 Connection: KF25 x 2
 Construction: 6 tier mesh pates, silicone insulating ring
 Recommended coolant  50% Ethylene glycol + 50% purified water
 Unit/shipping weight
 105 lbs / 145 lbs
 Unit/shipping dimensions 
 13 x 18 x 26" / 18 x 23 x 38" (WxDxH)
 Compliance  UL 471:2010 Ed.10+R:08Dec2016
 CSA C22.2#120 Issued: 2013/03/01 Ed: 4 Refrigeration Equipment
 Warranty  One year